11.4 Command Reference
────────────────────────────────────────────────────
Demand measurement stop time setting
Synta
x
DEMand:STOP:TIME data1,data2,data3,data4,data5
dat
a
data1 Year 1980 to 2079 (numerical data in NR1 format)
data2 Month 1 to 12 (numerical data in NR1 format)
data3 Day 1 to 31 (numerical data in NR1 format)
data4 Hours 0 to 23 (numerical data in NR1 format)
data5 Minutes 0 to 59 (numerical data in NR1 format)
Functio
n
Sets the demand measurement stop time.
Erro
r
If an attempt was made to specify a nonexistent date, a command error is
generated.
An execution error results if an attempt was made to set a time that
precedes the demand measurement start time.
In the following circumstances, a device dependent error is generated:
If an attempt was made to execute this command while waiting for or
during demand measurement.
In the event of a system error.
Exampl
e
Transmission
":DEM:STOP:TIME 1997,02,28,18,04"
Sets the demand measurement stop time to 16:18:04 on February 28, 1997.
Demand measurement stop time query
Synta
x
DEMand:STOP:TIME?
Functio
n
Returns the demand measurement stop time in (data1) through (data5), all
in the form of NR1 numeric data.
Respons
e
synta
x
If headers are ON
If headers are OFF
":DEMAND:STOP:TIME data1,data2,data3,data4,data5"
"data1,data2,data3,data4,data5"
Exampl
e
If headers are ON
If headers are OFF
Transmission
Response
":DEM:STOP:TIME?"
":DEMAND:STOP:TIME
1997,02,28,18,04"
":DEM:STOP:TIME?"
"1997,02,28,18,04"