EasyManua.ls Logo

Hioki 3166 - Page 209

Hioki 3166
380 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
187
────────────────────────────────────────────────────
11.4 Command Reference
────────────────────────────────────────────────────
Demand time setting
Synta
x
DEMand:TIME data
dat
a
0, 1, 2, 3, 4, 5, 6, 7, 8, 9, 10 (numerical data in NR1 format)
0: 5 minutes 6: 3 hours
1: 10 minutes 7: 4 hours
2: 15 minutes 8: 6 hours
3: 30 minutes 9: 8 hours
4: 1 hour 10: 12 hours
5: 2 hours
Functio
n
Sets the demand measurement time.
Erro
r
An execution error results if any value other than that indicated above is
set for (data).
In the following circumstances, a device dependent error is generated:
If an attempt was made to execute this command while waiting for or
during demand measurement.
In the event of a system error.
Exampl
e
Transmission
":DEM:TIME 3"
Sets the demand measurement time to 30 minutes.
Demand time query
Synta
x
DEMand:TIME?
Functio
n
Returns the demand period setting.
Respons
e
synta
x
If headers are ON
If headers are OFF
":DEMAND:TIME data"
"data"
Exampl
e
If headers are ON
If headers are OFF
Transmission
Response
":DEM:TIME?"
":DEMAND:TIME 3"
":DEM:TIME?"
"3"
D
EMand:TIME
D
EMand:TIME?

Table of Contents