188
────────────────────────────────────────────────────
11.4 Command Reference
────────────────────────────────────────────────────
■ Transformer capacity setting
Synta
DEMand:TRANs data
dat
0 to 1000000 (numerical data in NR1 format)
Functio
・Sets the transformer capacity. The unit is " VA".
・Although values in NRf format are accepted in (data).
Not
If the transformer capacity is set to "0", the TLF calculation is not made.
Erro
An execution error results if an attempt was made to set a value outside of
the range indicated above.
Exampl
Transmission
":DEM:TRAN 3000"
Sets the transformer capacity to 3 kVA.
■ Transformer capacity query
Synta
DEMand:TRANs?
Functio
Returns the transformer capacity setting by means of NR3 numeric data
(data).
Respons
synta
If headers are ON
If headers are OFF
":DEMAND data"
"data"
data
+1.2345E+03 (numerical data in NR3) format
Exampl
Transmission
Response
":DEM:TRAN?"
":DEMAND:TRANS +1.2345E+03"
":DEM:TRAN?"
"+1.2345E+03"
EMand:TRANs
EMand:TRANs?