11.4 Command Reference
────────────────────────────────────────────────────
Measurement line frequency setting
Synta
x
FREQuency data
dat
a
50, 60, 400 (numerical data in NR1 format)
Functio
n
Sets the frequency of the line that is to be measured.
Not
e
If this setting is not correct, it may be impossible to obtain correct
measurements.
Erro
r
An execution error results if any value other than that indicated above is
set for (data).
Exampl
e
Transmission
":FREQ 60"
Sets the frequency of the line being measured as "60 Hz".
Measurement line frequency setting query
Synta
x
FREQuency?
Functio
n
Returns the setting for the frequency of the line being measured.
Respons
e
synta
x
If headers are ON
If headers are OFF
":FREQUENCY data"
"data"
Exampl
e
If headers are ON
If headers are OFF
Transmission
Response
":FREQ?"
":FREQUENCY 60"
":FREQ?"
"60"
F
REQuency
F
REQuency?