11.4 Command Reference
────────────────────────────────────────────────────
■ Initial screen display
Synta
DISPlay:INItial
dat
No present
Functio
Displays the initial screen.
Not
If an attempt is made to execute this command while waiting for or
executing integrated measurement and demand measurement, a device-
dependent error results. Use the STOP command to release the power
meter from the integrating state before executing this command.
Erro
In the following circumstances, a device dependent error is generated:
If an attempt was made to execute this command while waiting for or
executing integrated measurement and demand measurement.
In the event of a system error.
Exampl
Transmission
":DISP:INI"
Displays the initial screen.
■ Data reset execution
Synta
DRESet
dat
No present
Functio
Executes a data reset, clearing the minimum/maximum value data,
integrated data, and demand data.
Erro
In the following circumstances, a device dependent error is generated:
If an attempt was made to execute this command while waiting for or
executing integrated measurement and demand measurement.
) In the event of a system error.
Exampl
Transmission
":DRES"
Executes a data reset.
ISPlay:INItial