11.4 Command Reference
────────────────────────────────────────────────────
■ Normal measurement mode display
Synta
DISPlay:NORMal data
dat
1, 2 (numerical data in NR1 format)
1:Instantaneous value
2: Minimum/maximum
Functio
Displays the Normal Measurement mode screen.
Not
・If this command is executed, the current maximum/minimum value data
and the integrated measurement data are reset.
・If an attempt was made to execute this command while waiting for or
executing integrated measurement or demand measurement. Use the
STOP command to release the power meter from the integrating state
before executing this command.
Erro
・In the following circumstances, a device dependent error is generated.
(1) If an attempt was made to execute this command while waiting for or
executing integrated measurement or demand measurement.
(2) In the event of a system error.
・An execution error results if any value other than that indicated above is
set for (data).
Exampl
Transmission
":DISP:NORM 1"
Displays the Normal Measurement mode "Instantaneous Value" screen.
ISPlay:NORMal