191
────────────────────────────────────────────────────
11.4 Command Reference
────────────────────────────────────────────────────
■ Integrated measurement mode display
Synta
DISPlay:INTEgrate data
dat
1, 2, 3, 4 (numerical data in NR1 format)
1: Instantaneous value
2: Minimum/maximum
3: Integrated 1/2
4: Integrated 2/2
Functio
Displays the Integrated Measurement mode screen.
Not
・If this command is executed, the current minimum/maximum value data
and the integrated measurement data are reset.
・If an attempt is made to execute this command while waiting for or during
demand measurement, a device-dependent error results. Use the STOP
command to release the power meter from the integrating state before
executing this command.
Erro
・In the following circumstances, a device dependent error is generate:
If an attempt was made to execute this command while waiting for or
during demand measurement.
In the event of a system error.
・An execution error results if any value other than that indicated above is
set for (data).
Exampl
Transmission
":DISP:INTE 1"
Displays the Integrated mode "Instantaneous Value" screen.
ISPlay:INTEgrate