204
────────────────────────────────────────────────────
11.4 Command Reference
────────────────────────────────────────────────────
■ Integrated measurement start method setting
Synta
INTEgrate:STARt:METHod data
dat
MANUAL, TIME (character data)
Functio
Sets the integrated measurement start method.
Erro
・In the following circumstances, a device dependent error is generated:
If an attempt was made to execute this command while waiting for or
during integrated measurement.
In the event of a system error.
・An command error results if an attempt was made to use a character that
cannot be used in (data).
Exampl
Transmission
":INTE:STAR:METH TIME"
Sets the integrated measurement start method to TIME.
■ Integrated measurement start method query
Synta
INTEgrate:STARt:METHod?
Functio
Returns the Integrated measurement start time settings in (data) in the
form of character data.
Respons
synta
If headers are ON
If headers are OFF
":INTEGRATE:START:METHOD data"
"data"
Exampl
Transmission
Response
":INTE:STAR:METH?"
":INTEGRATE:START:METHOD
TIME"
":INTE:STAR:METH?"
"TIME"
NTEgrate:STARt:METHod