11.4 Command Reference
────────────────────────────────────────────────────
Integrated measurement start in integrated measurement mode
Synta
x
INTEgrate:STARt
dat
a
No present
Functio
n
Starts integration.
Not
e
If the integration start method is set to "TIME", the power meter enters th
e
integrated measurement wait state. If the integrated measurement start
time has already passed when the power meter enters this state,
integration is not executed.
Erro
r
In the following circumstances, a device dependent error is generated:
If an attempt was made to execute this command while waiting for or
during integrated measurement.
In the event of a system error.
Exampl
e
Transmission
":INTE:STAR"
Starts integration (or puts the power meter in the integrated measurement
wait state).
I
NTEgrate:STARt