209
────────────────────────────────────────────────────
11.4 Command Reference
────────────────────────────────────────────────────
■ Integrated timer setting
Synta
INTEgrate:TIMEr data1,data2,data3
dat
data1 Hours 10 to 1000 (numerical data in NR1 format)
data2 Minutes 0 to 59 (numerical data in NR1 format)
data3 Seconds 0 to 59 (numerical data in NR1 format)
Functio
Sets the integrated timer (integration time).
Not
The minimum value for the integration timer is 10 seconds; the maximum
value is 1000 hours.
Erro
・An execution error results if an attempt is made to set a time outside of th
ranges indicated above.
・In the following circumstances, a device dependent error is generated:
If an attempt was made to execute this command while waiting for or
during integrated measurement.
In the event of a system error.
Exampl
Transmission
":INTE:TIME 10,0,0"
Sets the integration timer (integration time) to 10 hours.
■ Integrated timer query
Synta
INTEgrate:TIMEr?
Functio
Returns the integrated timer (integration time) setting.
Respons
synta
If headers are ON
If headers are OFF
":INTEGRATE:TIMER data1,data2,data3"
"data1,data2,data3"
Exampl
Transmission
Response
":INTE:TIME?"
":INTEGRATE:TIMER 10,0,0"
":INTE:TIME?"
"10,0,0"
NTEgrate:TIMEr
NTEgrate:TIMEr?