225
────────────────────────────────────────────────────
11.4 Command Reference
────────────────────────────────────────────────────
Reactive power operation method setting
Synta
x
OPERationvar data
dat
a
ON, OFF (character data)
ON: Calculates the reactive power by means of the reactive power meter
method.
OFF: Calculates the reactive power on the basis of the effective value.
Functio
n
Sets whether or not the reactive power meter method is to be used.
Not
e
If this command is executed, the current maximum/minimum value data is
reset. Does not utilize the reactive power meter method.
Erro
r
If data is set to character data other than "ON" or "OFF", a command error
is generated.
If an attempt was made to execute this command while waiting for or
executing integrated measurement and demand measurement.
Exampl
e
Transmission
":OPER OFF"
Specifies that the reactive meter method is not being used.
Reactive power operation method query
Synta
x
OPERationvar?
Functio
n
Returns the reactive power operation method setting query in (data)as
either ON or OFF.
Respons
e
synta
x
If headers are ON
If headers are OFF
":OPERATIONVAR data"
"data"
Exampl
e
If headers are ON
If headers are OFF
Transmission
Response
":OPER?"
":OPERATIONVAR OFF"
":OPER?"
"OFF"
O
PERationvar
O
PERationvar?