11.4 Command Reference
────────────────────────────────────────────────────
■ CT ratio setting
Synta
SCALe:CT data
dat
0.01 to 10000.00 (numerical data in NR2 format)
Functio
The current value of CT ratio is set.
Not
・If this setting is incorrect, correct measurements will not be obtained.
・If data is not in NRf format.
Erro
・If the value other than setting limit is set, an execution error is generated.
・If an attempt is made to execute this command while waiting for or
executing integrated measurement or demand measurement, a device-
dependent error results.
Exampl
Transmission
":SCAL:CT 1"
The CT ratio is set to 1.
■ CT ratio setting query
Synta
SCALe:CT?
Functio
The current value of CT ratio is returned as a numerical value (data)in
NR2 numerical format.
Not
With this query, if any error occurs, no response message is produced.
Respons
synta
If headers are ON
If headers are OFF
data
":SCALE:CT data"
"data"
+1.00000E‑02 to +1.00000E+04
(numerical data in NR3 format)
Exampl
Transmission
Response
":SCAL:CT?"
":SCALE:CT +1.00000E+00"
":SCAL:CT?"
"+1.00000E+00"
CALe:CT
CALe:CT?