229
────────────────────────────────────────────────────
11.4 Command Reference
────────────────────────────────────────────────────
■ PT ratio setting
Synta
SCALe:PT data
dat
1 to 10000 (numerical value in NR1 format)
Functio
・ The current value of PT ratio is set.
Not
・ If data is other than character data, a command error is generated.
・ Setting the averaging count.
Erro
・ If the value other than setting limit is set, an execution error is generated.
・ If an attempt is made to execute this command while waiting for or
executing integrated measurement or demand measurement, a device-
dependent error results.
Exampl
Transmission
":SCAL:PT 1"
The PT ratio is set to 1.
■ PT ratio setting query
Synta
SCALe:PT?
Functio
The current value of PT ratio is returned as a numerical value (data)in
NR1 numerical format.
Respons
synta
If headers are ON
If headers are OFF
":SCALE:PT data"
"data"
Exampl
Transmission
Response
":SCAL:PT?"
":SCALE:PT 1"
":SCAL:PT?"
"1"
CALe:PT