230
────────────────────────────────────────────────────
11.4 Command Reference
────────────────────────────────────────────────────
PLL synchronization source, frequency measurement source setting
Synta
x
SOURce data
dat
a
U1, I1 (character data)
Functio
n
Sets the source for PLL synchronization and for frequency measurement.
Not
e
If data is other than character data, a command error is generated.
Erro
r
If the value other than setting limit is set, an execution error is generated.
If an attempt is made to execute this command while waiting for or
executing integrated measurement or demand measurement, a device-
dependent error results.
Exampl
e
Transmission
":SOUR U1"
Sets the voltage (U1) as the source for PLL synchronization and for
frequency measurement.
PLL synchronization source, frequency measurement source query
Synta
x
SOURce?
Functio
n
Returns the PLL synchronization and frequency measurement source
setting in (data) as either U1 or I1.
Respons
e
synta
x
If headers are ON
If headers are OFF
":SOURCE data"
"data"
Exampl
e
If headers are ON
If headers are OFF
Transmission
Response
":SOUR?"
":SOURCE U1"
":SOUR?"
"U1"
S
OURce
S
OURce?