233
────────────────────────────────────────────────────
11.4 Command Reference
────────────────────────────────────────────────────
■ Voltage range setting value query
Synta
VOLTage?
Functio
Returns the presently set voltage range as a numerical value (data1)in
NR1 format, and also returns whether voltage auto-ranging is presently
taking place or not as "ON" or "OFF" (as data2).
Not
By using the "TRANsmit:SEParator" command, the message unit separator
can be changed from the semicolon ";" to the comma ",".
Respons
synta
If headers are ON
If headers are OFF
":VOLTAGE:RANGE data1;AUTO data2"
"data1;data2"
Exampl
Transmission
Response
":VOLT?"
":VOLTAGE:RANGE 150;AUTO
OFF"
":VOLT?"
"150;OFF"
OLTage?