288
────────────────────────────────────────────────────
16.3 Integrated Measurement Mode and Demand Measurement Mode
────────────────────────────────────────────────────
Indicates that power was lost durin
this interval
Printing Example After Power is Restored
4. Floppy disk operation
a. If power was lost while data was being saved to a floppy disk
The data that was being saved is not guaranteed. In the worst case, the file
itself may be damaged
If automatic output was being used when power was lost, a character string
indicating that power was lost is written to the floppy disk.
b. If power was lost while integrated measurement or demand measurement
was in progress
After power is restored, the floppy disk drive does nothing when power is
restored.
Reference The 3166 battery backup
・As a countermeasure against sudden loss of power, the 3166 CLAMP ON Hi
TESTER has a built-in battery that backs up memory containing the setting
conditions, minimum/maximum value data, and integrated data (including
demand measurements).
If the display data is being held, the battery backs up all of that data as
well.
・When the power is turned on, the setting conditions that were in effect the
last time that the power was turned off remain in effect.
・Because the battery backup maintains the integrated value data, it is
necessary to press the Data Reset key and clear the existing data in
"Integrated Measurement" mode if the results of the upcoming integrated
measurements are not to be added to the existing integrated value data.