EasyManua.ls Logo

Hioki IM3533 - Page 212

Hioki IM3533
400 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6.3 Judging with Upper and Lower Limit Values (Comparator Measurement Mode)
204
The set reference value and upper and lower limit values are common to percentage mode
and percentage deviation mode.
5
Settable range: -999.999% to 999.999%
Press and use the numeric keypad to
set the upper limit value.
Set the upper limit value as a percentage relative
to the reference value.
The actual internal operation consists of calculating the upper-limit value of comparison using the equation
given below, and comparing it to the calculated value to enable a decision to be made.
Upper limit comparison value = reference value + |reference value| ×
Upper Limit Value Setting
When you do not want to set the upper
limit: Press .
100
Percentage set value
6
Press to confirm the upper limit value.
The actual internal operation calculates the lower limit comparison value with the following equation, and when
a value that is lower than the reference value is set, the minus (-) sign is required for the percentage setting
value.
Lower limit comparison value = reference value + |reference value| ×
100
Percentage set value
Settable range: -999.999% to 999.999%
7
Return to step 2, press , use the numeric keypad to enter the lower limit value, and
press .
8
Press to close the setting screen.

Table of Contents

Related product manuals