4.2 Setting Basic Settings of Measurement Conditions
73
4
Chapter 4 LCR Function
The testing speed can be set. The slower the testing speed is, the more accurate are the results.
4.2.7 Setting the Measurement speed
The waveform averaging function allows you to set the measurement speed at a higher
level of detail. When the waveform averaging function is enabled, speed settings are not
available. Disable the waveform averaging function before setting the speed.
See "4.5.2 Setting the Detection Signal Waveform Averaging Count (Waveform Averaging
Function)" (p. 128)
Procedure
LCR Measurement Screen
1
LCR Basic Settings
Press .
2
LCR Basic Settings
To select the measurement speed.
3
Measurement speed varies with the measurement
conditions.
See "About Measurement Times and Measurement
Speed" (p. 350)
Performs high-speed measurement.
This is the normal measurement speed.
Measurement precision improves.
Measurement accuracy is better than
SLOW.
Measurement Speed Setting
4
Press to close the setting screen.