15.4 Discarding the Instrument
362
When disposing of this instrument, remove the lithium battery and dispose of battery and instrument in
accordance with local regulations.
15.4 Discarding the Instrument
To avoid electric shock, turn off the power switch and disconnect the power cord
and measurement cables before removing the lithium battery.
Battery may explode if mistreated. Do not short-circuit, recharge, disassemble or
dispose of in fire.
Keep batteries away from children to prevent accidental swallowing.
If the protective functions of the instrument are damaged, either remove it from service or
mark it clearly so that others do not use it inadvertently.
Lithium Battery Removal
Take care not to short the + and -.
Doing so may cause sparks.
CALIFORNIA, USA ONLY
This product contains a CR Coin Lithium Battery which
contains Perchlorate Material - special handling may apply.
See www.dtsc.ca.gov/hazardouswaste/perchlorate
Lithium Battery
(Overhead View)
1
Remove the six screws from the sides and
one screw from the rear.
2
Remove the cover.
Insert the tweezers between the battery and bat-
tery holder as shown in the diagram below and
lift up the battery.
3
Verify that the power is off, and remove the
connection cables and power cord.
4
Required tools:
One Philips screwdriver
One wire cutter (to remove the lithium battery)
Battery Holder