4.5 Setting Application Settings
137
4
Chapter 4 LCR Function
The higher the measurement frequency, the shorter the time that INDEX and EOM are high (off). When the
high (off) time is too short due to characteristics of the input circuit, the instrument can be configured to
maintain the low (on) state for a preset time once EOM
changes to low (on) before reverting the signal to high
(off) after the completion of measurement. The INDEX
output method can be changed in the same manner.
See "Chapter 12 External Control" (p. 309)
4.5.7 Setting the EOM Output Method
Procedure
LCR Measurement Screen
LCR Application Settings
1
Press .
2
LCR Application Settings
Setting the output method.
Use or to set the EOM
output time
for the PULSE setting.
For HOLD and PULSE timing charts, see
"Chapter 12 External Control" (p. 309).
3
Setting the EOM output method
• Settable range: 0.0001 s to 0.9999 s
• If you make a mistake during input:
press to cancel the input and start
again.
• The output time cannot be set unless the
output method has been set to [PULSE].
4
Press to close the setting screen.