14.1 General Specifications
337
14
Chapter 14 Specifications
Specifications Chapter 14
14.1 General Specifications
1.Basic Specifications
Measurement mode (1) LCR mode: Measurement with single condition
(2) ANALYZER mode (IM3533-01 only): Measurement frequency sweep
• No. of measurement points: 2 to 801
• Sweep method: Normal sweep START-STOP
• Display: List display
(3) TRANSFORMER mode: Measurement using a single condition
(4) CONTINUOUS measurement mode: Continuous measurement of saved conditions
• LCR mode.................. Up to 60
• ANALYZER mode ...... Up to 2 (IM3533-01 only)
* Continuous measurement using mixed LCR mode and ANALYZER mode operation is
also supported (IM3533-01 only).
Measurement items (1) LCR mode, ANALYZER mode:
Z (Impedance), Y (Admittance),
(Phase angle), Rs (Equivalent series resistance
(ESR)), Rp (Equivalent parallel resistance), X (Reactance), G (Conductance),
B (Susceptance), Ls (Equivalent series inductance), Lp (Equivalent parallel inductance),
Cs (Equivalent series capacitance), Cp (Equivalent parallel capacitance), Q (Q factor),
D (Loss coefficient tan, Rdc (DC resistance), T (Temperature)
(2) TRANSFORMER mode:
N (winding ratio), M (mutual inductance),
L (inductance difference)
Display range
Measurement frequency (1) Frequency range
1 mHz to 200 kHz
(2) Setting resolution
0.001 Hz to 99.999 Hz ........1 mHz steps
100.00 Hz to 999.99 Hz ......10 mHz steps
1.000 kHz to 9.9999 kHz.....100 mHz steps
10.000 kHz to 99.999 kHz...1 Hz steps
100.00 kHz to 200.00 kHz...10 Hz steps
(3) Frequency accuracy
±0.01% of setting or less
Parameters Display range (6 digit)
Z 0.00m to 9.99999G
Y 0.000n to 9.99999GS
±0.000 to 999.999
Rs, Rp, X, Rdc ±0.00m to 9.99999G
G, B ±0.000n to 9.99999GS
Cs, Cp ±0.0000p to 9.99999GF
Ls, Lp, M, L ±0.00000 to 9.99999GH
D ±0.00000 to 9.99999
Q ±0.00 to 9999.99
% ±0.000 to 999.999%
T -10.0 to +99.9
N 0.00000f to 999.999G