4.5 Setting Application Settings
138
You can save the measurement results inside the instrument (Up to 32,000 items).The saved measurement
results can be saved to a USB flash drive. They can also be acquired using a communication command. The
memory function setting is the same in mode, mode, and mode.
The items saved to memory are in accordance with the
:MEASure:VALid setting. For details on how to
acquire the saved measurement results or set
:MEASure:VALid, refer to LCR Application Disk (Commu-
nication Commands).
4.5.8 Saving Measurement Results (Memory function)
Saving Measurement Values
Procedure
LCR Initial Screen LCR Application Settings
1
Press .
LCR Application Settings
2
After touching to disable the mem-
ory function, use or to set the num-
ber of measurement results.
3
Settable range: 1 to 32000
The number of measurement results cannot be
changed while set to "OFF".
Memory Function Setting