4.3 Setting DC Resistance Measurement
85
4
Chapter 4 LCR Function
This delay time serves to delay measurement until offset measurement (0 VDC) stabilizes.
4.3.3 Setting the Offset Measurement Delay Time
(Adjustment Delay)
When number of averaging times is 1
Setting changed
• Changed from AC measurement to
DC measurement
• Range changed
Measure DC
resistance
DC delay
(Initial value: 0 seconds)
End measurementMeasure offset
ADJ delay
(Initial value: 0.0030
seconds)
When the number of averaging times is 2 or more
(The number of times is 2 in this example)
DC delay
(Initial value:
0 seconds)
Measure DC
resistance
Measure
offset
Measure
offset
Obtain
average value
Measurement for set number
of averaging times
End
measurement
Measurement for set number
of averaging times
Measure DC
resistance
Measure DC
resistance
DC delay
(Initial value:
0 seconds)
Measure offset
ADJ delay
(Initial value: 0.0030
seconds)
ADJ delay
(Initial value: 0.0030
seconds)
Setting changed
• Changed from AC measure-
ment to DC measurement
• Range changed
Procedure
LCR Measurement Screen
1
Rdc Basic Settings
2
Rdc Basic Settings
Press .