5.3 Application Settings
190
The higher the measurement frequency, the shorter the time that INDEX and EOM are high (off). When the
high (off) time is too short due to characteristics of the input circuit, the instrument can be configured to main-
tain the low (on) state for a preset time once EOM
changes to low (on) before reverting the signal to high (off)
after the completion of measurement. The INDEX
output method can be changed in the same manner.
See "Chapter 12 External Control" (p. 309)
5.3.11 Setting the EOM Output Method
Procedure
ANALYZER Application Settings
1
ANALYZER Measurement Screen
Press .
2
ANALYZER Application Settings
3
I/O Judgment Settings
Setting the output method.
Settable range: 0.0001 s to 0.9999 s
For HOLD and PULSE timing charts, see "Chapter
12 External Control" (p. 309).
Use or to set the EOM output time
for the PULSE setting.
4
Press to close the setting screen.