EasyManua.ls Logo

Hioki IM3533 - Displaying Average Values (Averaging Set)

Hioki IM3533
400 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4.2 Setting Basic Settings of Measurement Conditions
74
With the averaging function, the measured values can be averaged. Using this function, it is possible to
reduce fluctuations in the measured value display.
When the number of averaging times is 4, the number of measurements, measurement output points, and
measurement value calculation method during output are as follows.
4.2.8 Displaying Average Values (Averaging Set)
With internal trigger
A rolling average of the tested values over the set number of times for
averaging is always calculated backwards from the present.
(When the sample to be tested is changed over, it takes a little time for a certain stabili-
zation time period until the results is reliable.)
With external trigger
Average over the number of averaging times based on trigger input.
M1 M2 M3 M4 M5 M6
M7 M8 M9 M10
M1
1
M1+M2
2
M1+M2+M3
3
M1+M2+M3+M4
4
M2+M3+M4+M5
4
M3+M4+M5+M6
4
M4+M5+M6+M7
4
M5+M6+M7+M8
4
M6+M7+M8+M9
4
M7+M8+M9+M10
4
Measurement points
Measurement value output
Moving average
Time
M1 M2 M3 M4 M5 M6 M7 M8 M9 M10
Time
Arithmetic mean
M1+M2+M3+M4
4
M5+M6+M7+M8
4
Measurement points
Measurement value output