2.2 Pre-Operation Inspection
30
Please read the "Follow these precautions to ensure safe operation and to obtain the
full benefits of the various functions." (p. 5) before use.
2.2 Pre-Operation Inspection
Before using the instrument for the first time, verify that it operates normally to ensure that no damage
occurred during storage or shipping. If you find any damage, contact your authorized Hioki distributor or
reseller.
Do not use the instrument if damage
is found, as electric shock or short-
circuit accidents could result. Contact
your authorized Hioki distributor or
reseller.
Metal Exposed
Is the power cord insulation torn, or is
any metal exposed?
1
No Metal Exposed
Peripheral Device Inspection
Is the insulation on a measurement
cable torn, or is any metal exposed?
Metal Exposed
If there is any damage, measured val-
ues may be unstable and measure-
ment errors may occur. Replace the
cable with an undamaged one.
No Metal Exposed
If damage is evident, request repairs.
Yes
Instrument Inspection
When turning power on
No
2
The power cord may be damaged, or
the instrument may be damaged in-
ternally. Request repairs.
No
Yes
The instrument may be damaged
internally. Request repairs.
See "15.1 Inspection, Repair and
Cleaning" (p. 353)
"15.3 Error display" (p. 360)
An error indication
occurs (ERR)
No
Inspection complete
Is damage to the instrument evident?
Is there an error display on the Opening
screen?
Does the Opening screen appear (mod-
el no., version no.)?
Opening screen