8.3 Compensating Values to Match Reference Values (Load Compensation)
233
8
Chapter 8 Error Compensation
Be sure to set the cable length before performing load compensation.
See"8.4 Compensating Measurement Cable Errors (Cable Length Compensation)" (p. 245)
Load compensation is enabled when the current measurement conditions match all com-
pensation conditions.
If the current measurement frequency and compensation frequency do not match, an
error such as the following will be displayed on the Measurement screen, and compensa-
tion will be canceled.
If a condition other than the compensation frequency does not match, compensation is
performed but an error like the following is displayed in the measurement screen.
When the OPEN or SHORT compensation is valid, the load compensation is performed
for Z and processed by the OPEN or SHORT compensation.
In acquiring load compensation data (i.e., reference sample measurement), the OPEN/
SHORT compensation settings, that were defined before entry into the Load Compensa-
tion Screen, are valid.
If the setting of the low Z high accuracy mode is changed, the compensation value
becomes invalid.