8.3 Compensating Values to Match Reference Values (Load Compensation)
233
8
Chapter 8 Error Compensation
• Be sure to set the cable length before performing load compensation.
See"8.4 Compensating Measurement Cable Errors (Cable Length Compensation)" (p. 245)
• Load compensation is enabled when the current measurement conditions match all com-
pensation conditions.
• If the current measurement frequency and compensation frequency do not match, an
error such as the following will be displayed on the Measurement screen, and compensa-
tion will be canceled.
• If a condition other than the compensation frequency does not match, compensation is
performed but an error like the following is displayed in the measurement screen.
• When the OPEN or SHORT compensation is valid, the load compensation is performed
for Z and processed by the OPEN or SHORT compensation.
• In acquiring load compensation data (i.e., reference sample measurement), the OPEN/
SHORT compensation settings, that were defined before entry into the Load Compensa-
tion Screen, are valid.
• If the setting of the low Z high accuracy mode is changed, the compensation value
becomes invalid.