Setting the Output Conditions
138
ZW-7000/5000
User's Manual
Setting the Average Filter
Output measured data as the average value of a preset count.
This is set in cases such as to reduce the fluctuation of measurement values and increase the static resolution.
Multi View Explore : [Bank] | [(Bank Data Name)] (double click)
Edit pane : [Task Settings] icon ( )
Task Settings window : [Filter]
When measurement is started, the measurement result is not updated and output until the measured values have
been sampled for the preset average count.
Item Setting item Setting value Description
Average number
of times
Average Internal synchronous measurement mode
1/2/4/8/16/32/64/128/256/512/1024(default value)/2048/4096/
8192/16384 [
times]
Switches the average count.
External/PDO synchronous measurement mode
1(default value)/2/4/8/16/32/64/128/256/512/1024/2048/4096/
8192/16384 [
times]
1
Select how many times the latest measured
values are to be averaged before output from
[Average].
The average filter can also be set by the operating keys on the Sensor Controller.
Setting the Average Filter p.218