Channel Menu
Operating Manual 1145.1084.12 – 30 282
Triggered Meas. Sequence
Selects the measurement cycle or sequence of actions to be triggered.
Sweep means that each trigger event starts an entire sweep, according to the current sweep
configuration.
Sweep Segment means that each trigger event starts the next sweep segment within the current
sweep. If a sweep type other than Segmented Sweep is active, this option is equivalent to Sweep.
Point means that each trigger event starts the measurement at the next sweep point.
Partial Measurement means that each trigger event starts the next partial measurement at the
current or at the next sweep point. If each measurement points requires only one partial
measurement , this option is equivalent to Point.
See also background information in section Alternating Sweeps.
Relation to other sweep settings
Some sweep settings are logically incompatible with a particular selection of the Triggered Meas.
Sequence:
If a Time sweep is performed, the Triggered Meas. Sequence is always a sweep.
Alternating sweep mode only differs from normal sweep mode if the measurement sequence
comprises more than one sweep point and more than one partial measurement. If Point is
selected, Alternating Sweeps has no effect.
The period of the trigger events should be adjusted to the triggered measurement sequence. If the
analyzer receives a trigger event while the last sequence is still running, the trigger event is skipped and a
message Trigger event missed is displayed.
Trigger Delay
Specifies a delay time between the trigger event and the start of the measurement. If the triggered
measurement sequence is a Partial Measurement, it is possible to define source port-specific trigger
delays. For an overview of trigger timing scenarios see background information in section Alternating
Sweeps.