Remote Control Commands
R&S
®
SMBV100A
680Operating Manual 1176.8016.02 ─ 17
[:SOURce<hw>]:BB:DM:DLISt:DATA:APPend <Bits>
The command appends the bit data onto the end of the existing data in the selected
data list. This means that existing content in the data list is not overwritten. By this
means very long data lists can be built up piecemeal. The data format is as specified in
command SOURce:BB:DM:DLISt:DATA.
The command cannot be used with an empty data list, such as one that has just been
created, for example. In this case the command SOURce:BB:DM:DLISt:DATA must
first be used to enter modulation data in the list.
*RST has no effect on data lists.
Setting parameters:
<Bits> 0 | 1 {,0 | 1 } | block data
Example:
BB:DM:DLIS:SEL ' d_list2'
selects data list d_list2.
FORM ASC
selects ASCII data transmission format.
BB:DM:DLIS:DATA:APP 1,1,1,0,0,0,1,1,0,1...
adds the specified numeric data to the existing data in data list
d_list2.
Usage: Setting only
Manual operation: See "Data List Editor" on page 364
[:SOURce<hw>]:BB:DM:DLISt:DELete <Filename>
The command deletes the specified data list. from the default directory. The default
directory is set using command MMEM:CDIRectory. A path can also be specified, in
which case the files in the specified directory are deleted. The file extension may be
omitted. Only files with the file extension *.dm_iqd will be deleted.
Setting parameters:
<Filename> string
Example:
BB:DM:DLIS:DEL ' d_list2'
deletes data list d_list2.
Usage: Setting only
Manual operation: See "File Manager" on page 363
[:SOURce<hw>]:BB:DM:DLISt:POINts <Points>
The command defines the number of bits in the selected data list to be utilized. When a
list is being filled with block data, this data is only ever sent in multiples of 8 bits. How-
ever the exact number of bits to be exploited can be set to a different figure. The
superfluous bits in the list are then ignored.
SOURce Subsystem