CURSor:VBArs:DELTa?
Returns the time or frequency difference between the two vertical bar cursors.
The units (seconds or Hertz) are specified by the CURSor:VBArs:UNIts
command. If the cursor source is an FFT math waveform,
CURSor:VBArs:DELTa is always in Hertz, regardless of the value set by
CURSor:VBArs:UNIts. Query only.
NOTE. If Trigger View is active, this query returns 9.9E37 and generates event
221 (Settings conflict).
Group
Cursor
Syntax
CURSor:VBArs:DELTa?
Returns
<NR3>
Examples
CURSor:VBArs:DELTa? might return 8.92E-1, indicating that the time
difference between the vertical bar cursors is 0.892 seconds.
CURSor:VBArs:HPOS<x>?
Returns the horizontal value of the specified vertical bar ticks for cursor <x>. The
units are specified by the CURSor:HBArs:UNIts query. <x> specifies the cursor.
Valid values are 1 and 2. Query only.
Group
Cursor
C commands
84 TBS2000 Series Programmer