RECAll:WAVEForm
Recalls a saved waveform file to a reference location. Command only, no query
form.
Group
Save and Recall
Syntax
RECAll:WAVEForm <file path>,REF<x>
Related Commands
SAVe:WAVEform on page 210, FILESystem:CWD on page 126, FILESystem? on
page 125
Arguments
<file path> specifies a location of the instrument setup file. <file path> is a
quoted string that defines the file name and path. Input the file path using the
form <drive>:/<dir>/<filename>.<extension> and one or more <dir>s are
optional. If you do not specify them, the instrument will read the file from the
default directory (see FILESystem:CWD). <filename> stands for a filename; the
use of wildcard characters in filenames is not supported. Filename extensions are
not required, but highly recommended.
REF<x> is the instrument reference memory location in which to load the
waveform. You must load a saved waveform into a reference memory location
before displaying the waveform. Reference memory location values are 1 or 2.
Examples
RECALL:WAVEFORM "tek00000.isf",REF1 recalls the waveform stored in the
file named tek00000.isf from the current directory for waveforms to the reference
location 1.
R commands
TBS2000 Series Programmer 199

Table of Contents

Other manuals for Tektronix TBS2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TBS2000 Series and is the answer not in the manual?

Tektronix TBS2000 Series Specifications

General IconGeneral
BrandTektronix
ModelTBS2000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals