WFMOutpre:YMUlt?
Returns the vertical scale factor per digitizing level in units specified by
WFMOutpre:YUNit for the waveform specified by the Returns the vertical scale
factor per digitizing level in units specified by WFMOutpre:YUNit for the
waveform specified by the DATa:SOUrce command. The query command will
time out and an error is generated if the waveform specified by DATa:SOUrce is
not turned on. command. The query command will time out and an error is
generated if the waveform specified by DATa:SOUrce is not turned on. (Query
Only)
Group
Waveform
Syntax
WFMOutpre:YMUlt?
Related commands
DATa:SOUrce on page 91, WFMInpre:YMUlt on page 254
Examples
WFMOUTPRE:YMULT? might return :WFMOUTPRE:YMULT 4.0000E-3
indicating that the vertical scale for the corresponding waveform is 100 mV/div
(for 8-bit waveform data).
WFMOutpre:YOFf?
Returns the vertical position in digitizing levels for the waveform specified by the
DATa:SOUrce command. The query command will time out and an error will be
generated if the waveform specified by DATa:SOUrce is not turned on. Query
only.
Group
Waveform
W commands
TBS2000 Series Programmer 265