TRIGger:A:LEVel:CH<x>
Sets or queries the trigger level for the specified channel. Each channel can have
an independent level.
Group
Trigger
Syntax
TRIGger:A:LEVel:CH<x> {<NR3>|TTL|ECL}
TRIGger:A:LEVel:CH<x>?
Arguments
ECL specifies a preset ECL high level of -1.3V.
TTL specifies a preset TTL high level of 1.4V.
<NR3> specifies the trigger level in user units (usually volts).
Examples
TRIGGER:A:LEVEL:CH3 TTL sets the A edge trigger to TTL high level for
channel 3.
TRIGGER:A:LEVEL:CH2? might return :TRIGGER:A:LEVEL:CH2 1.3000E
+00 indicating that the A edge trigger is set to 1.3 V for channel 2.
TRIGger:A:LOWerthreshold:CH<x>
Sets or queries the lower threshold for the channel selected. Each channel can
have an independent level. Used in Runt trigger as the lower threshold. Used for
all other trigger types as the single level/threshold.