EasyManua.ls Logo

Tektronix TBS2000 Series

Tektronix TBS2000 Series
328 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CH<x>?
Returns the vertical parameters for the specified channel. The value of <x> can
vary from 1 through 4 depending on instrument model. Query only.
Because CH<x>:SCAle and CH<x>:VOLts are identical, only CH<x>:SCAle is
returned.
Group
Vertical
Syntax
CH<x>?
Related Commands
SELect:REF<x> on page 216
Returns
instrument vertical settings for the specified channel.
Examples
CH1? might return :CH1:SCALE 1.0E0;POSITION 0.0E0; COUPLING
DC;BANDWIDTH FULL;PROBE 1.0E0.
CH<x>:BANdwidth
Sets or queries the selectable low-pass bandwidth limit filter setting of the
specified instrument channel. The value of <x> can vary from 1 through
4 depending on instrument model.
This command is equivalent to setting the BW Limit option in the Vertical menu.
Group
Vertical
C commands
60 TBS2000 Series Programmer

Table of Contents

Other manuals for Tektronix TBS2000 Series

Related product manuals