CH<x>?
Returns the vertical parameters for the specified channel. The value of <x> can
vary from 1 through 4 depending on instrument model. Query only.
Because CH<x>:SCAle and CH<x>:VOLts are identical, only CH<x>:SCAle is
returned.
Group
Vertical
Syntax
CH<x>?
Related Commands
SELect:REF<x> on page 216
Returns
instrument vertical settings for the specified channel.
Examples
CH1? might return :CH1:SCALE 1.0E0;POSITION 0.0E0; COUPLING
DC;BANDWIDTH FULL;PROBE 1.0E0.
CH<x>:BANdwidth
Sets or queries the selectable low-pass bandwidth limit filter setting of the
specified instrument channel. The value of <x> can vary from 1 through
4 depending on instrument model.
This command is equivalent to setting the BW Limit option in the Vertical menu.
Group
Vertical
C commands
60 TBS2000 Series Programmer