EasyManua.ls Logo

Tektronix TBS2000 Series

Tektronix TBS2000 Series
328 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
WFMOutpre:WFId?
Returns a string describing several aspects of the acquisition parameters for the
waveform specified by the DATa:SOUrce command. The query command will
time out and an error will be generated if the waveform specified by
DATa:SOUrce is not turned on. Query only.
Group
Waveform
Syntax
WFMOutpre:WFId?
Related commands
DATa:SOUrce on page 91
Returns
<QString> comprises the following comma-separated fields:
Source The source identification string as it appears in the front-panel scale factor
readouts.
Coupling A string describing the vertical coupling of the waveform.
Vert Scale A string containing the vertical scale factor of the unzoomed
waveform. The numeric portion will always be four digits. The examples cover
all known internal units.
Horiz Scale A string containing the horizontal scale factor of the unzoomed
waveform. The numeric portion will always be four digits. The examples cover
all known internal units.
Record Length A string containing the number of waveform points available in
the entire record. The numeric portion is given as an integer.
Acquisition Mode A string describing the mode used to acquire the waveform.
Examples
WFMOUTPRE:WFID? might return :WFMOUTPRE:WFID "Ch1, DC coupling,
100.0mVolts/div,500.0μs/div, 1000 points, Sample mode".
W commands
262 TBS2000 Series Programmer

Table of Contents

Other manuals for Tektronix TBS2000 Series

Related product manuals