CH<x>:PRObe:GAIN
Sets or queries the gain factor for the probe attached to the channel specified by
<x>, where x is the channel number. The gain of a probe is the output divided by
the input transfer ratio. For example, a common 10x probe has a gain of 0.1.
Group
Vertical
Syntax
CH<x>:PRObe:GAIN <NR3>
CH<x>:PRObe:GAIN?
Related commands
CH<x>:SCAle on page 72
Arguments
<NR3> is the probe gain. Allowed values depend on the specific probe.
Examples
CH1:PRObe:GAIN 0.1 sets the channel 1 probe gain to 0.1.
CH2:PROBE:GAIN? might return :CH2:PROBE:GAIN 0.1000E+00 indicating
that the attached 10x probe delivers 1 V to the channel 2 BNC for every 10 V
applied to the probe input.
C commands