WFMOutpre:XINcr?
Returns the horizontal point spacing in units of WFMOutpre:XUNit for the
waveform specified by the DATa:SOUrce command. This value corresponds to
the sampling interval. The query command will time out and an error will be
generated if the waveform specified by DATa:SOUrce is not turned on. Query
only.
Group
Waveform
Syntax
WFMOutpre:XINcr?
Related commands
DATa:SOUrce on page 91, WFMOutpre:XUNit? on page 263
Examples
WFMOUTPRE:XINCR? might return :WFMOUTPRE:XINCR 10.0000E-6
indicating that the horizontal sampling interval is 10 μs/point.
WFMOutpre:XUNit?
Returns the horizontal units for the waveform specified by the DATa:SOUrce
command. The query command will time out and an error will be generated if the
waveform specified by DATa:SOUrce is not turned on. Query only.
Group
Waveform
Syntax
WFMOutpre:XUNit?