*RST
(Reset) Returns the instrument to a known set of instrument settings, but does not
purge any stored settings. This command executes a subset of the FACtory
command.
The *RST command does not change the following items:
Calibration data that affects device specifications
Output queue
Service Request Enable Register settings
Power-On Status Clear flag setting
Alias definitions
Stored settings or waveforms
The *PUD? response
Any of the values associated with the DATA command
Instrument password
Group
Status and Error
Syntax
*RST
Related Commands
FACtory on page 119, *PSC on page 195, *RCL on page 197, RECAll:SETUp on
page 198, *SAV on page 205, SAVe:SETUp on page 209
Arguments
None
Examples
*RST resets the instrument settings to factory defaults.
R commands
204 TBS2000 Series Programmer