CH<x>:PRObe:DEGAUss:STATE?
Returns the state of the probe degauss for the channel specified by <x>, where is
x is the channel number. Query Only.
NOTE. This command will return PASSED for probes that do not support degauss
operations.
Group
Vertical
Syntax
CH<x>:PRObe:DEGAUss:STATE?
Returns
NEEDED indicates the probe should be degaussed before taking measurements.
RECOMMENDED indicates the measurement accuracy might be improved by
degaussing the probe.
PASSED indicates the probe is degaussed.
FAILED indicates the degauss operation failed.
RUNNING indicates the probe degauss operation is currently in progress.
Examples
CH1:PRObe:DEGAUss:STATE? might return :CH1:PRObe:DEGAUss:STATE
FAILED indicating the degauss operation failed.