TRIGger:FREQuency?
Returns the edge or pulse width trigger frequency. This is the same as the readout
in the lower right corner of the screen. Query only.
Group
Trigger
Syntax
TRIGger:FREQuency?
Returns
Edge or pulse width trigger frequency.
Examples
TRIGger:FREQuency? might return TRIGGER:FREQUENCY 1.0E3.
TRIGger:STATE?
Returns the current state of the triggering system. Query only.
Group
Trigger
Syntax
TRIGger:STATE?
Related commands
TRIGger:A:MODe on page 230
Returns
ARMED indicates that the instrument is acquiring pretrigger information. All
triggers are ignored when TRIGger:STATE is ARMed.
AUTO indicates that the instrument is in the automatic mode and acquires data
even in the absence of a trigger.
READY indicates that all pretrigger information has been acquired and that the
instrument is ready to accept a trigger.
SAVE indicates that the instrument is in save mode and is not acquiring data.
TRIGGER indicates that the instrument triggered and is acquiring the post trigger
information.
Examples
TRIGGER:STATE? might return :TRIGGER:STATE ARMED indicating that
the pretrigger data is being acquired.
T commands