Page #60 background imageLoading...
Page #60 background image
Group
Acquisition
Syntax
ACQuire:STATE {OFF|ON|RUN|STOP|<NR1>}
ACQuire:STATE?
Related Commands
*OPC on page 193, ACQuire:STOPAfter on page 42
Arguments
OFF|STOP|<NR1> = 0 stops acquisitions; any other value starts acquisitions..
ON|RUN|<NR1> ≠ 0 starts acquisition and display of waveforms.
Examples
ACQuire:STATE RUN starts acquisition of waveform data and resets the number
of acquisitions count (NUMACq) to zero.
ACQuire:STATE? might return: ACQUIRE:STATE 0 indicating that the
acquisition is stopped.
ACQuire:STOPAfter
Sets or returns whether the instrument continually acquires acquisitions or
acquires a single sequence.
Group
Acquisition
Syntax
ACQuire:STOPAfter {RUNSTop|SEQuence}
ACQuire:STOPAfter?
A commands
42 TBS2000 Series Programmer

Table of Contents

Other manuals for Tektronix TBS2000 Series