SAVe:ASSIgn:TYPe
Sets or queries the assignment of the data to be saved when the front-panel Save
button is pressed.
Group
Save and Recall
Syntax
SAVe:ASSIgn:TYPe {IMAGe|WAVEform|SETUp|PRINTer}
SAVe:ASSIgn:TYPe?
Arguments
IMAGe assigns the Save button to save screen images.
WAVEform assigns the Save button to save waveforms.
SETUp assigns the Save button to save setups.
PRINTer assigns the Save button to printer.
Examples
SAVe:ASSIgn:TYPe WAVEform sets the data to be saved to waveform.
SAVe:ASSIgn:TYPe? might return :SAVe:ASSIgn:TYPe WAVEform indicating
that a wavefrom will be saved when the Save button is pressed.
SAVe:IMAge
Saves the screen image to a file. Command only, no query form.
Supported image formats are png, windows bitmap, and jpg. The format to use is
determined by the value obtained from the :SAVe:IMAge:FILEFormat? query.
Group
Save and Recall
S commands
206 TBS2000 Series Programmer