MEASUrement:MEAS<x>:SOUrce1
Sets or queries the source for all single source measurements and specifies the
source to measure from when taking a delay measurement or phase measurement.
Where <x> identifies the measurement, 1 through 6.
This is equivalent to selecting the measurement source in the MEASURE menu.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:SOUrce1 {CH<x>|MATH}
MEASUrement:MEAS<x>:SOUrce1?
Arguments
CH<x> specifies the input channel source for the measurement.
MATH specifies the measurement source channel as the math waveform.
Examples
MEASUREMENT:MEAS2:SOURCE1 CH1 sets source1 for Measurement 2 to
channel 1.
MEASUrement:MEAS1:SOUrce1? might
return :MEASUrement:MEAS1:SOUrce1 MATH indicating the source for
measurement 1 is the math waveform.