CALibrate:FACtory
Starts the instrument internal factory calibration operation. The calibration
operation consists of a sequence of steps. You send the CALibrate:CONTINUE
command to advance to the next calibration step. The calibration program
automatically sets up the instrument for each calibration step. Use the
CALibrate:ABOrt command to abort the factory calibration. Command only, no
query form.
NOTE. You should only use this command in a qualified service environment. For
more information about the factory calibration sequence, refer to the service
manual for your instrument.
You can only send synchronization commands or queries (such as *OPC, OPC,
*WAI, BUSY) while doing a factory calibration.
Group
Calibration and Diagnostic
Syntax
CALibrate:FACtory {STARt|CONTinue|PREVious|ABOrt|DUmp}
Arguments
STARt initializes the factory calibration sequence and starts the first calibration
step.
CONTinue begins the next factory calibration step.
PREVious attempts to run the most recent factory calibration step again.
ABOrt stops the calibration process.
DUmp stops the calibration and prints the calibration constants.
Examples
CALibrate:FACtory STARt starts the factory calibration process.
C commands
54 TBS2000 Series Programmer

Table of Contents

Other manuals for Tektronix TBS2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TBS2000 Series and is the answer not in the manual?

Tektronix TBS2000 Series Specifications

General IconGeneral
BrandTektronix
ModelTBS2000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals