CALibrate:FACtory
Starts the instrument internal factory calibration operation. The calibration
operation consists of a sequence of steps. You send the CALibrate:CONTINUE
command to advance to the next calibration step. The calibration program
automatically sets up the instrument for each calibration step. Use the
CALibrate:ABOrt command to abort the factory calibration. Command only, no
query form.
NOTE. You should only use this command in a qualified service environment. For
more information about the factory calibration sequence, refer to the service
manual for your instrument.
You can only send synchronization commands or queries (such as *OPC, OPC,
*WAI, BUSY) while doing a factory calibration.
Group
Calibration and Diagnostic
Syntax
CALibrate:FACtory {STARt|CONTinue|PREVious|ABOrt|DUmp}
Arguments
STARt initializes the factory calibration sequence and starts the first calibration
step.
CONTinue begins the next factory calibration step.
PREVious attempts to run the most recent factory calibration step again.
ABOrt stops the calibration process.
DUmp stops the calibration and prints the calibration constants.
Examples
CALibrate:FACtory STARt starts the factory calibration process.