HORizontal:RECOrdlength:Auto
Sets or queries the horizontal record length mode.
Group
Horizontal
Syntax
HORizontal:RECOrdlength:Auto <NR1>
HORizontal:RECOrdlength:Auto?
Arguments
<NR1> sets the record length mode. 1 enables auto record length mode,
0 disables auto record length mode.
Examples
HORizontal:RECOrdlength:Auto 1 enables auto record length mode of the
analog channels.
HORizontal:RECOrdlength:Auto? might
return :HORizontal:RECOrdlength:Auto 0 indicating that auto record length
mode of the analog channels is off.
HORizontal:RESOlution
Sets or returns the horizontal record length of acquired waveforms. The sample
rate is automatically adjusted at the same time to maintain a constant time per
division. The query form of this command returns the current horizontal record
length.
Group
Horizontal
Syntax
HORizontal:RESOlution <NR1>
HORizontal:RESOlution?
H commands