Examples
CURSor:HBArs:UNIts? might return :CURSOR:HBARS:UNITS BASE
indicating that the units for the horizontal bar cursors are base.
CURSor:HBArs:USE
Sets the horizontal bar cursor measurement scale. This command is only
applicable when ratio cursors are on. No query form.
Group
Cursor
Syntax
CURSor:HBArs:USE {CURrent|HALFgrat|FIVEdivs}
Related commands
CURSor:HBArs:UNIts on page 80
Arguments
CURrent sets the H Bar measurement scale so that 0% is the current position of
the lowest H Bar cursor and 100% is the current position of the highest H Bar
cursor.
HALFgrat resets the H bar measurement scale to half the number of divisions
(five for some models and four for others) so that 25% is the current position of
the lowest H Bar cursor and 75% is the current position of the highest H Bar.
FIVEdivs sets H Bar measurement scale so that five screen major divisions is
100%, where 0% is –2.5 divisions and 100% is +2.5 divisions from the center
horizontal graticule.
Examples
CURSOR:HBARS:USE FIVEDIVS sets the H Bar measurement scale so that
5 screen major divisions equals 100%.
C commands
TBS2000 Series Programmer 81