SAVe:SETUp
Saves the current state of the instrument into the specified memory location. This
is equivalent to selecting the Save Setup option in the Save/Recall Setup menu.
You can later use the *RCL command to restore the instrument to this saved
state. Command only, no query form.
Group
Save and Recall
Syntax
SAVe:SETUp {<file path>|<NR1>}
Related Commands
RECAll:SETUp on page 198, *RCL on page 197, *SAV on page 205
Arguments
<NR1> specifies a location for saving the current front-panel setup. The front-
panel setup value ranges from 1 to 10. Using an out-of-range value causes an
execution error. Any settings that have been stored previously at this location will
be overwritten.
<file path> is the target location for storing the setup file. <file path> is a quoted
string that defines the file name and path. Input the file path using the form
<drive>:<dir>/<filename>. <extension> and one or more <dir>s are optional. If
you do not specify them, the instrument will store the file in the current working
directory. <filename> stands for a filename. (Use of wildcard characters in
filenames is not supported.) Filename extensions are not required but are highly
recommended. For setups, use the extension .SET.
Examples
SAVe:SETUp5 saves the current front-panel setup to memory location 5.
SAVE:SETUP "TEK00000.SET" saves the current instrument setup in the file
TEK00000.SET in the current working directory.
S commands
TBS2000 Series Programmer 209