DIAg:SELect:<function>
Runs self-tests on the specified system subsystem. No query form.
Group
Calibration and diagnostic
Syntax
DIAg:SELect:<function>
Arguments
<function> specifies a single instrument subsystem on which to run self tests
(diagnostics). Valid values are:
■
ACQ tests the acquisition system.
■
CPU tests the CPU.
■
DISplay tests the display.
■
FPAnel tests the front panel controls.
■
IO tests the IO ports.
■
ROM tests the system read only memory.
Examples
DIAg:SELect:ACQ specifies to run self tests on the acquisition system.
DIAg:STATE
This command starts or stops the instrument self-test. Depending on the
argument, self-test capabilities are either turned on or off. No query form.
Group
Calibration and diagnostic
Syntax
DIAg:STATE {EXECute|ABORt}
D commands