Examples
MEASUREMENT:MEAS3:UNITS? might
return :MEASUREMENT:MEAS1:UNIts % indicating units for measurement
1 are set to percent.
MEASUrement:MEAS<x>:VALue?
Returns the value that was calculated for the instrument on-screen periodic
measurement specified by <x>. Where <x> identifies the measurement, 1 through
6. Query only.
This is the same value as displayed on-screen. If measurement statistics are
enabled, a new value is calculated with every waveform. In addition, this value is
updated approximately every 1/3 second. If you are acquiring a long acquisition
record, the instrument may take longer to update.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:VALue?
Related commands
MEASUrement:MEAS<x>:UNIts? on page 189, *ESR? on page 107, ALLEv? on
page 48
Returns
<NR3> is the measurement value.
Examples
MEASUREMENT:MEAS3:VALUE? might return 28.75E6 if measurement
number three is frequency. If the measurement has an error or warning associated
with it, then an item is added to the error queue. The error can be checked for
with the *ESR? and ALLEv? commands.
M commands
190 TBS2000 Series Programmer

Table of Contents

Other manuals for Tektronix TBS2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TBS2000 Series and is the answer not in the manual?

Tektronix TBS2000 Series Specifications

General IconGeneral
BrandTektronix
ModelTBS2000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals